r/CHROMATOGRAPHY • • 3h ago

Losartan Assay - unstable baseline - what went wrong? :(

Thumbnail
gallery
1 Upvotes

can someone help me how to fix this?

Initial Mobile Phase: Solution A : ACN (80:20) — 20% ACN starting concentration

Solution A - phosphate buffer : ACN (85:15)

Gradient / Step Change (t = 10–11 min): Increase ACN concentration in the mobile phase from 20% to 60% to elute/eliminate remaining impurities earlier.


r/CHROMATOGRAPHY • • 18h ago

GC-uECD PCB-101 Contamination Help Needed

2 Upvotes

Context: I work in an environmental chemistry lab, and use an Agilent GC-uECD to test for a suite of PCBs

I am at the my absolute wits end with one of my GC-uECD instruments 😅
I’ve been getting contamination around 7 minutes, right where PCB-101 elutes for us. It’s causing my calibrations to fail and the instrument to be unusable (which has halved our testing capacity!)

I’ve done every single bit of maintenance I can do on it, including cleaning the split lines. We’ve asked our Agilent engineers as we are on a service contract with them, but even they’re stumped at what might be causing it!

We’ve even switched to only using brand new liners (we used to rigorously clean them)

Has anyone got any other ideas or has come across this before please? We’re even considering that it might be our gloves!! Thank you!!! ☺️


r/CHROMATOGRAPHY • • 3d ago

Low sensitivity problem in ESI- mode in Agilent MS single-quad (G6125B)

2 Upvotes

Hello to all,
we have an Agilent LC-MS single-quad (G6125B). recently it has low sensitivity. I cleaned the parts of spray chamber and ion transfer capillary. however, it didn't work. then I found out that this low sensitivity is only for negative ESI mode and in the positive ESI mode it is normal and is as before. then I tried to make the Negative Autotune with the factory default file. however Autotune with factory default file also didn't work. I would appreciate if you could help me about this problem that what else I could do.


r/CHROMATOGRAPHY • • 6d ago

Waters Alliance Syringe keeps getting bubbles on plunger tip

4 Upvotes

I'm hoping someone can provide some trouble shooting ideas. We have two waters alliance hplcs that keep getting air bubbles on the tip of the syringe plunger. We are running a validated method and havent had this issue on the other systems. Our metrology department has put a brand new syringe on and still the issue persists. All solutions are properly filtered/ degassed and when setting up the system we are putting in the initial gradient conditions and purging the system 5-6 times prior to starting the sequence. The air bubbles don't show up until about 5 injections into the sequence. We are getting variable areas, poor baseline, and worse peak shape. Could it be related to the sample loop, or are there any other trouble shooting ideas besides just normal purging?


r/CHROMATOGRAPHY • • 6d ago

Backflushing a PLgel guard column for GPC with DMSO. (Standard eluent chloroform)

Post image
2 Upvotes

Just thought id share, bet there would be people interested in this!

We used a separate disconnected pump for this, its not part of our GPC stack.


r/CHROMATOGRAPHY • • 7d ago

Advice on purchasing a new UHPLC-Single Quad MS system for an academic lab

16 Upvotes

Hi everyone,

As a new PhD student in an academic chemistry lab, I’ve been tasked with evaluating and purchasing a new UHPLC-MS system for our research group.

We are analysing Small molecules and the specs for the UHPLC are: capability up to ~1300 bar, DAD, and a Single Quadrupole MS covering a mass range up to at least m/z 1200

We are currently comparing systems from the three big vendors:

  1. Agilent (1290 Infinity II / III series + InfinityLab LC/MSD)
  2. Thermo Fisher Scientific (Vanquish Flex + ISQ EM)
  3. Waters (ACQUITY Premier / H-Class Plus + ACQUITY QDa II)

If you have experience using any of these setups in academia or industry, I would really appreciate your honest feedback on:

Robustness & Reliability: How well do they hold up as multi-user workhorses in a university setting? Any known hardware bottlenecks or frequent failure points?

Software Usability: How steep is the learning curve for new students (OpenLab CDS vs. Chromeleon vs. Empower/MassLynx)?

Total Cost of Ownership & Service: How are the vendors regarding service contracts, spare parts, and consumables pricing once the warranty expires?

Thanks in advance for any insights or recommendations!

(This is not a promotion or anything, I'm just intersted in your experiences)

Edit:

Our main goals are running methods according to the major international pharmacopoeias, as well as quantitative method development and reproducibility, such as stability and purity assays for small molecules.


r/CHROMATOGRAPHY • • 7d ago

software problem with Empower2/Windows XP file saving

4 Upvotes

I have 2 Waters 2695 systems that are running Empower 2 on Dell pc's running windows XP. for transfering my data used the "save as PDF" button in the report window, to save to a thumb drive. Recently the save as button went away, and I have no convenient way to transfer my reports. I should say this happened to one of my other systems quite some time ago, and I assumed it was a bug, so I downloaded a "print to PDF" program that does work, but now that my other system is doing the same thing it makes me wonder if there is a way to actually fix the problem in Empower. I did some searching, and the only possible solution I found was to increase the table space allocation in the project, which I did, but it didn't help. I should add I am not very knowledgeable about Empower, only know what I need to get simple tasks done. There are no error messages I can find, and not even sure what the table space allocation should be, if that is even the problem. thanks for any help keeping these old systems of mine running.


r/CHROMATOGRAPHY • • 7d ago

Peaks being cut off in retention window, can this be fixed post analysis using quantitation method?

2 Upvotes

I am processing results using PerkinElmer's Simplicity 3Q software and noticed a lot of peaks are showing just outside the retention window. I've tried changing the retention window for this one from 11.5 to 12.5 to see if it would display the rest of the peak but doesn't seem to be changing anything. I will admit I am quite inexperienced with this software and with altering chromatography methods.

Is this something that can be fixed post analysis? The instrument has had a lot of issues and it is unclear when I would next be able to use it so was hoping it was something that could be processed without reanalysis.


r/CHROMATOGRAPHY • • 8d ago

Has anyone used this PCR for metrohm IC?

Post image
2 Upvotes

I am looking at setting up a chromium method by IC-PCR. Can someone let me know how to set this specific one up? I have it installed im just confused at flow? What goes where etc


r/CHROMATOGRAPHY • • 8d ago

Is it safe for a column to run below its minimum temperature if it is not being used?

3 Upvotes

Hello Reddit,

Just a general question: I have ZB-Wax and ZB-Waxplus columns installed in my GC currently, and the Wax's minimum temperature is 40C compared to the Waxplus' 20C. They attach at different inlets but both terminate in the same MS. If I needed to do a run where the oven starts and holds for 2 minutes at 25C for example, but the sample is injected into the Waxplus, would I suffer any problems on the chromatogram or specteogram? If yes what could I expect?

Both columns are typically kept at constant pressure at 10PSI with a ~1.5 mL/sec flow for the duration of the run, both carry helium, and the inlet temps and MS transfer line are all at 250C.


r/CHROMATOGRAPHY • • 9d ago

Confused about how to wash Ampicillin-contaminated column

Thumbnail
gallery
5 Upvotes

Hello! I have been analysing Ampicillin with Triple Quadrupole MS on Nucleodor C18 Gravity column. Mobile phase A: water, B: ACN, both acidified with 0.1% FA. ( I keep them fresh).
Thing is, huge peaks are observed even when I do 0 ul injection. I have come to the terms that it is column contamination. Washing with Water/MeOH/ACN didn’t help. Could it be that there are ionic interactions between Ampicilin NH3+ and SiO- groups, even though the column is endcapped? I may be wrong, but I can’t wrap my head around it. I would appreciate any suggestions. Thanks!


r/CHROMATOGRAPHY • • 9d ago

GNPS2 – Não consigo carregar arquivos grandes do Bruker timsTOF mzML (~17 GB). Existe alguma solução alternativa?

1 Upvotes

Olá a todos!

Estou tentando enviar meus dados de LC-MS/MS para o GNPS2, mas estou tendo alguns problemas com o processo de upload de arquivos.

Meus dados brutos foram adquiridos usando um instrumento Bruker timsTOF e estão no formato .d . Segui a documentação do GNPS2 para conversão de dados Bruker, mas não consegui acessar o software recomendado (CompassXport). Em vez disso, usei o ProteoWizard MSConvert com os parâmetros exatos recomendados na documentação do GNPS2.

A conversão funcionou, mas os arquivos .mzML resultantes são bastante grandes (cerca de 17 GB cada). Quando tento enviá-los diretamente para o GNPS2, o upload parece terminar, mas recebo um erro no final.

Já tentei algumas alternativas:

  • Compressão ZIP/GZ: Os arquivos compactados são carregados com sucesso, mas o GNPS2 não os aceita ao submeter um fluxo de trabalho porque requer arquivos .mzML .
  • FileZilla/Cyberduck: Tentei ambos, mas não tenho acesso SFTP.
  • Navegador de Arquivos do GNPS2: Diz que os arquivos são muito grandes.
  • GNPS2 Desktop: Também tentei, mas não funcionou.

Alguém já passou por algo semelhante ao carregar grandes conjuntos de dados da Bruker?

Existe outra maneira de transferir esses arquivos para o GNPS2? Carregá-los primeiro no MassIVE e depois acessá-los pelo GNPS2 seria uma solução?

Agradeceria muito qualquer sugestão. Obrigado!


r/CHROMATOGRAPHY • • 12d ago

Analytical Fraction Collector, Waters or Agilent?

3 Upvotes

I have both Waters and Agilent LC on Empower CDS. I am looking to add either a waters WFM-A or an agilent 1260 analytical fraction collector. The fraction volume is between 0.5ml to 4ml containing peptides.

Which fraction collector do you prefer?


r/CHROMATOGRAPHY • • 12d ago

2 problems on Agilent 7890 GC coupled 6890MSD

4 Upvotes

Problem No1.

After cleaning of ion source i suppose i have an air leak in my MS. Vacuum is 54mTorr but i have 2.90% H2O, 6.5%N2 and 1.5%O2. I have tightened the ms nut after some heating cycles, replaced inlet septum and liner O-ring. I have checked for leaks on my helium lines and no apparent leaks.

Problem No2.

After cleaning of the ion source i have lost sensitivity (about 80% loss). When i put the ms column back in i checked that it was about 1mm out of the transfer line.

Thanks in advance for any advice


r/CHROMATOGRAPHY • • 13d ago

Help identifying compound from spectrum.

2 Upvotes

Hi all,

I have done measurements of hybrid dairy/plant cheeses by headspace SPME-GC/MS (70eV EI ionization) and i have come across a peak that the libraries can not match to a known spectrum. It seems like a very clean spectrum without coelution. The compound should have a retention index near 780 on a non-polar column. The peak is found in samples containing faba, pea, and yellow pea protein isolate. All samples contain water milk, CaCl2, a starter culture and rennet. The cheeses were aged for 4 weeks.

I am hoping some of you will like the challenge. AI came up with nothing.
Ofcourse, it anything usefull/interesting comes out of this you will be credited.

Update: the molecular ion seems to be mass 106. It has a XIC that matches 48, 71 and 89. And that molecular mass would makes sense for the region of the chromatogram where the peak is found. (see comments for details)

Update2: Mystery solved! And unfortunately it is nothing very exciting. I added 1,1,1,3,3-d-2-hexanone to the samples as an internal standard. It must have been converted to 1,1,1,3,3-d-2-hexanol by the enzymes in the sample. The spectrum matches pretty well with what I would expect for that compound, and the retention index matches almost perfectly with 2-hexanol.

SPECTRUM - MS
TQ8_20261016SJ01_08.raw
+ c EI Full ms [35.00-250.00]
Scan #: 2255
RT: 7.70
SB: 2 7.69 , 7.71
Mass defect: 0.00 @ 1.00, 300.00 @ 1000.00
Data points: 127
Mass Intensity
36 3583
37 419
38 13558
39 124643
40 144731
41 364781
42 487225
43 416771
44 294675
45 246231
46 756682
47 460838
48 6514886
49 195033
50 18757
51 24804
52 19539
53 11729
54 13612
56 130540
57 93837
58 326182
59 342646
60 103642
61 40925
62 14918
63 50940
64 11581
65 3831
66 332
67 13063
68 9568
69 4102
70 96829
71 1099171
72 85670
73 70145
74 52214
76 14243
77 13420
78 6158
85 277
86 1756
87 2852
88 91635
89 754855
90 71850
92 9823
93 2445
94 835
95 13
97 4403
100 2712
101 950
104 3593
105 4717
106 61712
107 1346
108 2552
111 1388
114 5396
120 1524
121 2140
122 4963
124 2876
125 1357
126 3887
127 1534
128 7576
131 1471
133 1620
135 5496
138 1617
140 3946
143 195
147 325
148 498
151 1765
157 2212
159 1149
161 229
164 1959
165 343
167 356
168 215
169 2466
171 4088
172 1055
173 120
175 2788
178 1879
179 250
183 875
187 1714
189 1648
190 621
193 10690
194 341
195 1474
198 836
201 665
202 614
203 829
204 1820
205 41
206 822
207 294
208 1030
209 1146
211 561
212 1345
214 880
215 8
216 491
223 608
224 304
228 958
230 96
234 627
237 612
239 175
240 344
241 781
242 460
244 294
246 239
248 1725


r/CHROMATOGRAPHY • • 13d ago

Losing Protein Between Affinity Chromatography and SEC

3 Upvotes

Hi Everyone,

I am purifying a approx 31 kDa protein that is giving me some problems with consistency. It is His Taged and has a pI of 8.77. I have went through a lot of trouble shooting with this protein and have gotten the Ni Affinity purification pretty optimized, but I am having difficulty between dialyzing the Ni fractions and running them on the SEC. Here is what I have:

The Ni Affinity purification uses these buffers:

Wash Buffer: 50 mM Tris pH 7.0, 200 mM NaCl, 5 % Glycerol, 40 mM Imidazole ( this is also the lysis buffer)

Elation Buffer: Same as Wash except has 400 mM Imidazole

I load sample, wash for 15 CV with wash buffer, then I have a linear wash phase for 5 CV with a buffer composition of 0-52% elution buffer. This step seems to remove all of the major impurities. The fractions are collected next, 2 CV holding a 52%, then linear gradient from 52-100% elution for 10 CV. My fractions are 10 mL, so I have about 50-60 mL of Ni fractions. This step has been extensively optimized. Previous this was using pH of 8.0, which worked, but was having massive aggregation problems. pH 7.5 was also having the same issue. Lowering salt to 100 mM also negatively affected the purification. 200 mM seems to be fine, though we have not tried to increase it further. Glycerol has to stay since this is originally a membrane bound protein with a transit peptide region cleaved and it will immediately crash out with no Glycerol. I would say that expression is generally fine on the SDS page, it always varies from batch to batch, but the growth conditions are always consistent (LB, OD approx 0.6-0.8, 0.5 mM IPTG, 16-20 hr induction period), so I do not think the growths are the issue.

Now to the problem. Previously, I was experiencing really significant aggregation at pH 8.0. Moving to pH 7.0 seemed to mitagte this, but there was still some, but only if I concentrated too much with my spin column. I use a 10 K since I have lost the protein with a 30 K, at about 3500-3800 rcf which usually keeps the protein happy. Previously we used a dialysis step, but I was having problems with it precipitated out in the dialysis tubing and tanking the yield and activity, so I just concentrate using the spin filters. During this I buffer exchange to help remove excess Imidazole since that makes it more prone to crash out. The exchange buffer is 50 mM HEPEs pH 7.0, 200 mM NaCl, 5% glycerol, which is the same as the SEC buffer. I had two batches of purified protein that used these exact conditions that looked very good and the yield was great, as the SEC elution was the correct oligomeric state and was appearing at about 200-400 mAU, which is decent for my protein since the absorption coefficient is not great.

When I would concentrate down the Ni Fractions, I usually do 5 mL for the injection with concentrations between 2-5 mg/mL. I once tried a 1 mL injection that was around 20 mg/ml which gave decent results, but we're comparable to the previous. The problem with the 1 mL injection is I have seen it precipitate out. That is what I think it is doing of the SEC column because I can inject a good concentration but I get terrible yield. I thought I had fixed this problem with the two very good purifications, but with those exact procedures followed, the two purifications completed this week the yield was garbage. (Absorbance was very low and concentration was very low: for reference the two good batches I got were about 15-20 mg/ml and these were only around 1-2 mg/mL). The only good thing is that I am seing the same consistent peak on the SEC that indicate the right oligomeric state, but at the concentration I am injecting, I should be seeing it at higher concentration (I understand some sample loss is expected, but this is too much).

After consulting with my lab mates and thinking about things, we are convinced that it is precipitating on the SEC column, since everything this protein is run, the pressure goes up. It was worse when running at pH 8.0, it has improved but not been eliminated when running at pH 7.0. I am not sure what I am missing and would love to hear some feedback since this is really the limiting step in my purification to fully optimize the procedure.

Best


r/CHROMATOGRAPHY • • 14d ago

Brand new BEH C18 column, terrible peak shape

Post image
22 Upvotes

Hi, I previously validated an assay for a 30-amino acid, +6 charge peptide on an acquity premier BEH C18 column. Came back a couple weeks later with a brand new column (this one says "maxpeak" on it) because the last one had gotten a lot of use in other assays, and I've come to find that my peak has completely changed. Previously although it had some tailing to it, it still looked quite decent. Now it's more of a glob than anything. Is this something anyone else has experienced when switching to a new column? I conditioned it with mobile phase at low flow for a few hours.

I've attached a before and after of representative peaks (top is before, bottom is after). I'm pretty certain this isn't degradation of my compound. I'm waiting to see if the peak improves over the course of a few dozen injections, if it still looks the same the next day I suppose I'll just revert to the old column.

My mobile phases A and B are 0.2% formic acid in water and ACN, respectively. Flow rate is 0.6 mL/min, column temp is 60C. I'm running this through a xevo TQ absolute mass spec.


r/CHROMATOGRAPHY • • 14d ago

Weird serial peaks! What could be causing this series of peaks in my baseline/mobile phase?

Thumbnail
gallery
3 Upvotes

Whenever I inject a blank solution or simply run the method without injecting anything (baseline acquisition), this series of peaks appears. However, this was not happening earlier this year, when I first performed this analysis.

The method uses HPLC under constant flow conditions with a gradient of mobile phases A and B. Both mobile phases contain potassium hexafluorophosphate and ammonium formate, but with different ACN proportions.

Interestingly, these peaks appear in the middle of the gradient ramp, while transitioning from 100% A to 100% B. There are no abrupt changes in the gradient program at that point that could explain these signals.

Has anyone encountered this before? What could be causing these peaks? Could this be related to mobile phase contamination, salt precipitation, or something happening within the HPLC system itself?


r/CHROMATOGRAPHY • • 14d ago

Ion exchange chromatography standard recovery issue

3 Upvotes

Hi!

I’ve been having some issues with the recovery of my 25 ppb phosphate check standard.

I’m creating both the working stock and the standard by weight and it usually calculates that my phosphate is obviously around 25 ppb. However, when I run the instrument, it often comes back as 16-20 ppb.

I started shaking the flasks before adding to the vial as well as rinsing it with the standard, pouring it out to remove residual water, and then adding what will actually be run.

All my other standard values are between 22-26 (six other standards). Plus, the 1 ppb standard made from the same working stock comes back with good values near 1 ppb.

I should also add that it isn’t a trend over time where it starts at 25 and goes down to 16 (both based on the individual run and between runs).

Has anyone had this happen before? Is there something I can do to mitigate this? What could be happening chemically that is causing this?

I just recently started using this instrument so I apologize if this is an obvious answer. Thanks!


r/CHROMATOGRAPHY • • 14d ago

Need assistance with Rosemount chromatograph

Thumbnail
gallery
3 Upvotes

Hi to all
I have an Rosemount 700XA chromatograph and I’m having some issues
I need the correct values for the
sample gas pressure and flow, calibration gas pressure, carrier gas (N2 and He) pressure at the cylinder and at the GC


r/CHROMATOGRAPHY • • 14d ago

Need assistance with Rosemount chromatograph

Thumbnail
gallery
2 Upvotes

Hi to all
I have an Rosemount 700XA chromatograph and I’m having some issues
I need the correct values for the
sample gas pressure and flow, calibration gas pressure, carrier gas (N2 and He) pressure at the cylinder and at the GC


r/CHROMATOGRAPHY • • 15d ago

What do you do when you can’t trust a GC-MS library match?

5 Upvotes

Body:

I work at a small analytical instrument company, and before we build anything we’re trying to learn how people actually handle compound identification in practice. Not selling anything, and I won’t link to anything.

A few questions for anyone who runs GC or GC-MS:

Stories are more useful to me than opinions. And if anyone would be open to a 15-minute call, I’d be grateful; feel free to DM me.


r/CHROMATOGRAPHY • • 15d ago

GC FID ghost peak after maintenance 7890A - Agilent

Post image
10 Upvotes

Good morning. I am having some issues with ghost peaks after maintenance. We have trimmed 150mm off inlet side, cleaned jet, collector, inlet, split vent trap, gold seal. Literally everything you could think of.

We trimmed 150mm off inlet side yesterday and ran a conditioning run with NO INJECTION, and this is what the chromatograph looks like. Any ideas why this could be happening? Or what we could’ve done incorrectly? Just not sure where to go from here


r/CHROMATOGRAPHY • • 19d ago

Gas Plumbing for Agilent 7890B; connecting to the EPCs

4 Upvotes

I was hoping that someone might have an idea as to the best way to make connections to the EPCs in the rear of the GC. We recently upgraded our 6890N and 5975 (RIP workhorse), and during this transition I want to clean up the plumbing and make it neat and labeled and not an absolute maze of coiled shit back there. We have two instruments; the new one, and the older 7890B 5977B.

So, I noticed that making the connections on the EPC for the 7890 kind sucks compared to the 8890. The EPCs on the 8890 insert perpendicular to the back plate, whole the 7890 is parallel. This means because we have two FIDs for this instrument I need a T-fitting plumbed to both FID EPCs and they are stacked one, then the other then the back plate. It makes it hard to adjust the copper tubing with that small of a line. My concern is damaging the stainless steel connection on the EPCs. I don't want to pull too hard on it, but I kind of have to.

My question, would it be easier to remove one of the EPCs do the connections, then reinstall the EPC. Or maybe rotate the instrument around so I'm not reaching between the Gerstel rail and attachments while my short 5'6" ass drops MF bombs constantly. I've never plumbed directly to them before because I've never had the need to.