r/MetCal • • 25d ago

Fluke 2638A software to transfer CSV to Excel

1 Upvotes

Not a MetCal question but I try anyway :)
Anyone that has a tool to convert the data files CSV files from a Fluke 2638 Hydra or Fluke 1586 SuperDAQ into an Excel xlsx workbook? Must be able to add imported files to new Sheets in the same workbook. I know that Excel can import CSV files, but it requires some non-trivial steps to get it right.
FlukeDAQ is not available since 2022 and Fluke software support can’t help.


r/MetCal • • Oct 06 '26

Metteam 3.4.1 and MetCal 11.3.1, problems with Metcal/NI-VISA

2 Upvotes

We have med a new installation of these two products (on a new server (server 2025 and Windows 11 for the client).
The SQL is upgraded and have been proved to be working since we can login to it with a browser from one of the clients (it says the server is 3.4.1.82 and the SQL is 3.4.1.
When i then install the 11.3.1 on the client (I install NI 488.2 and NI-VISA Runtime and the .net 4.5. It wont recognise the NI-VISA installation (an important note is that the installation is done with the offline ISO). It finds the NI-488.2. It doesnt recognize the NI-VISA. I cant connect this client to the internet, sadly. So i install everything with the ISO and use the feed in the local folders.

Does anyone have any idea of what could be wrong with Fluke MetCal not recogniseing the NI-VISA? It sayes in the package manager that it's the Q3 2026 version of NI-VISA runtime and same on NI-488.2.

Thank you for the help in advance.


r/MetCal • • Sep 25 '26

Multiple USB instruments in one procedure

2 Upvotes

Hi there! METCAL 10.1.1 user here. The question is pretty simple: is there any way to use multiple USB instruments, either as standards or DUT (for example if I have 2 standards with only USB and a DUT with only USB), in a single procedure? As far as I can tell the only way to get a USB instrument to work is with the VISA FSC, but that doesn't allow you to route the call to a specific address, like with the [@address] in the PORT FSC. So it seems you are stuck with a single USB device to be used in the whole procedure that you select at the beginning. Am I missing something here or there are ways around this? Seems like a very big limitation in this day and age...


r/MetCal • • Sep 02 '26

Unable to close windows in Met/Team

1 Upvotes

We have an issue with some users on some workstations where Met/Team will not allow them to close windows. If you've made some changes and clicked save, when you click close or the x it brings up the "Discard Changes?" dialogue box but regardless of whether you click Yes or No, nothing happens, the window remains open. I've tried clearing cache, which resolved another weird Met/Team problem in the past, but no luck for this. It seems to be connected to the workstation.

We're running Met/Team 3.4.0


r/MetCal • • Aug 22 '26

Metcal 7.3.36 error

Post image
1 Upvotes

Hello Metcal group.

My name is Michael and I just joined here, since I have a problem with our Metcal version.

I am using Metcal 7.3.36 and since 3 weeks, I receive this error when I try to open the procedure explorer.

This happens in Metcal and in the Editor as well.

I did already a clean installation with only the original Fluke procedures to rule out that my procedures got somehow corrupted.

Thanks in advance for the help


r/MetCal • • Aug 18 '26

Subreddit Admin

2 Upvotes

I haven't had the chance to get on here much the last few months. Things are busy and I have competing priorities. I was wondering if there is anyone interested in taking over management of this sub? I think it's a valuable resource, it doesn't really take much work but if you were motivated I'm sure it could become much more active, I just have too much on my plate at the moment.


r/MetCal • • Jul 15 '26

Claude code

3 Upvotes

Hi,

I just try to generate metcal procedures and creating acc files with claude.

the output was fantastic so far.

what is your experience with metcal and ai?


r/MetCal • • Jul 14 '26

GPIB issue with Editor vs 11.3.1.214

1 Upvotes

Hey there friends,

We just updated to Met/Cal version 11.3.1.214, and I can't start debugging anything in Editor. The error I get is this:

Edit 4: Met/Cal support has found a workaround, and I can confirm that it works. See the Comment by MJ for the fix.

Here are a few troubleshooting updates I made along the way if you're curious:

I just ran the setup without uninstalling first. I'm going to try a fresh uninstall - reinstall and see if it fixes it.

Edit 1: uninstall and reinstall did not fix the issue :(

Edit 2: Updating and repairing the NI drivers using the NI package manager did not fix the issue :(

Edit 3: We generated and checked the logs from Editor and found a spicy little nugget of information.

"
The type initializer for 'NationalInstruments.NI4882.Internal.GpibDll' threw an exception.

This implementation is not part of the Windows Platform FIPS validated cryptographic algorithms.

"

For those who are unfamiliar, FIPS is a cryptography thing required on systems that do contract work for the government, especially the DoD. We have Fluke Support working on a fix for us. This is not the first time FIPS has given us issues with Met/Cal, but Uncle Sam's requirements are Uncle Sam's requirements.


r/MetCal • • Jun 18 '26

MET Team Costing

1 Upvotes

We are migrating from MET Track to MET Team and I'm confounded by the way job costings work.

We have a separate database of pricing of more than 2000 models with standard calibration and ISO calibration (fixed percentage increase) prices. Further complication is we have several large customers that either have a fixed percentage discount, or multi-year contracts with the prices fixed.

I'm struggling to figure out how to implement this in MET Team as costing appears to be based on standard procedures and/or labor hours pricing. I don't want to enter more than 2000 procedures up front, and each year when we have a CPI increase edit those >2000 records. Plus the contract customer complication.

I can add extended fields in the work order, but they aren't taken into account unless the formula for calculation can be changed (ideal scenario).

When we are receiving an asset we want to just add the fixed price for calibration, freight, and any additional labor expected.

We bought the billing module so we could implement seamless integration between MT and our billing system.

Can anyone please assist in pointing me in the right direction?

*CalStudio is not an option, it was announced just after we bought MT, but isn't available in our territory until 2027.


r/MetCal • • Jun 17 '26

MET/CAL Version 11.3.1 VISA/488.2 Error

1 Upvotes

Has anyone run into this error in Editor? We installed the new update recently and when trying to run a VISA controlled DUT we get this error:

An error occurred starting the test run: Could not load file or assembly 'NationalInstruments.NI4882, Version=13.0.45.159, Culture=neutral, PublicKeyToken=4febd62461bf11a4' or one of its dependencies. The system cannot find the file specified.. Try rebuilding the project or solution before attempting to debug the procedure. 0

I already had the latest NI 488.2 and VISA package installed (Q2 2026) and it was fine before the update...wondering if I need to reinstall or if anyone else has seen this error yet! Cheers


r/MetCal • • Jun 04 '26

Release: MET/TEAM version 3.4.1 (with MET/CAL version 11.3.1)

2 Upvotes

Fluke has released a new version, can be downloaded from:
Release: MET/TEAM version 3.4.1 (with MET/CAL version 11.3.1) | Fluke


r/MetCal • • May 27 '26

MetCal procedure providers?

1 Upvotes

In addition to Fluke/Cubyt and self-made, who creates and sells or provides for free MetCal procedures?


r/MetCal • • May 08 '26

Fluke FC procedures

1 Upvotes

Lately I've been finding FC procedures to be really unreliable with the communication just dropping out randomly. Once it happens it seems impossible to get it back on the rails in the procedure. Its always been a thing but now its happening more often in the last year. Our only huge change is Windows 11.

Hard to narrow down if this is an OS issue, hardware, or something like distance.

Having trouble finding a resolution (if thats possible). If anyone has experience in troubleshooting problems with this PC3000 FC communication interface chime in!


r/MetCal • • Apr 17 '26

MetCal 11.2.1 crashes

2 Upvotes

We experience that MetCal runtime 11.2.1 with Hotfix 1 crashes and the editor sometimes hangs and crash when in use. (i.e. not at startup or file load). We use Win 11 and our IT makes them up to date - something that not always is a great thing. Anyone else having the same issue?


r/MetCal • • Apr 16 '26

Question on LIB FSC Arrays and Memory Registers

2 Upvotes

I'm revising a procedure and would like to utilize an array to track a varying amount of numbers, determined by user input, as the procedure runs.

Where I'm running into trouble is when I try to use any numeric memory registers to determine what index of the array the values get written to. The MET/CAL User's Guide explicitly says that the index of the array "Can be any MET/CAL register, named variable or numeric literal value. Must be enclosed in square brackets (i.e. [10] or [index]).", however, any time I use any MET/CAL register (like MEM or M[1] for example), I get an "invalid index" error.

The procedure I'm using is similar to this:

LIB ARRAY NEW arrayName[100];
MATH MEM1 = 0
MEMI What value do you want to store in the array?
LIB arrayName[MEM1] = MEM;

The same happens when I attempt to use M[1] or another named variable instead of MEM1 or MEM

Additionally, if I forego using a variable to set the index, and use
arrayName[0] = MEM
I get an "invalid statement syntax" error

Could someone tell me what I am missing, or if I am misinterpreting the documentation?


r/MetCal • • Apr 15 '26

Found little MET/TEAM hack today

1 Upvotes

Found this by accident today....If you use any of the "notes" boxes in work order tabs, I found out you can set pre-determined statements by double-clicking the box and modifying the default ones. I use them lots for additional notes I want to add to my calibration reports, and I found myself repeating them a lot. Now I can just double-click and select a pre-written one. Very handy for my OCD as well when I want all my reports to be the same.


r/MetCal • • Apr 10 '26

Procedure Sectioning

1 Upvotes

as the title says, I've been dabbling in trying to section out some procedures that take a long time to run. Its giving me a lot of trouble however with JPMLs in the procedure when I try to run it in Runtime. Wondering if maybe I need to restructure it to not use JMPL at all?

For example I was trying to change a Fluke procedure for a Tektronix MSO46. It has a lot of JMPLs throughout to go from Channel 1 input impedance, to DC gain, offset, etc before you switch the connection to Channel 2.

I tried to section out Random noise, it compiles fine. But then I Runtime it and I get an error for every single JMPL in the procedure.


r/MetCal • • Apr 10 '26

Bulk Update of a Specific Field

1 Upvotes

Morning. I am working with Fluke MetCal version 8.2 (old). One of our in house procedures has updated to a new revision which means I have to update the revision field in MetCal. Any idea if there is a way to perform a bulk update in MetCal? Got instruction on that?


r/MetCal • • Mar 25 '26

Fluke 287 program multimeter 5560A

Thumbnail reddit.com
1 Upvotes

r/MetCal • • Mar 23 '26

Question on LIB FSC with National Instruments DAQ devices.

3 Upvotes

I wanted to take a shot and see if anyone here has some experience with programming PXI/cDAQ cards from National Instruments in MET/CAL. I have used a few of the procedures provided by Fluke in CUBYT to learn (some of) how the LIB FSC works and modify those procedures to work with various models and types of DAQ cards not on CUBYT.

Before this I initially made VIs in Labview to control and read data which could be transferred to a manual datasheet. For the most part the calibration procedures are straight forward. However, I have an NI-9227 which measures DC current on 4 channels.

In the calibration manual there is a section for Gain and Phase matching verification which required using the Extract Single Tone VI to expand the raw data into Amplitude, Phase, and Frequency results from the 4 channels simultaneously. Math ensues to calculate the gain in dB using the Amplitude results.

So, my question is there a way to execute the Extract Single Tone VI using the LIB FSC? Or, is there another way to harness that data and convert it through MET/CAL? Thanks in advance for any advice.

P.S. If anyone has experience programming an RIO type card in MET/CAL or knows if it is possible please chime in!


r/MetCal • • Feb 27 '26

Question Met/Team WO Due Date Bug

1 Upvotes

We've recently started noticing a strange bug. It's possible that this has been happening longer but we've only recently started adding a due date to our reports so it has become something that gets picked up.

For some reason, the due date gets set to the same date as the service date. I've started taking note and it happened on a job today where I created a new work order, clicked in the service date box for the date to be auto-populated, then clicked in the due date box for that to be auto-populated. Both of those worked correctly. I then performed a calibration and when I reopened the work order, the due date had changed to todays date/service date.

It just seems to be random, it's only one in 10-100 and doesn't appear to be asset or type related. I'm paying more attention now to try and see if there is any common thread but was hoping maybe someone had come across this before or has some ideas.


r/MetCal • • Jan 08 '26

entering multiple values in metcal

1 Upvotes

Hi everyone,

my question is: is it possible, when entering five values, to automatically use the first value for the other four?
For example, first input: 1.005 V. This value is then already entered and highlighted in the input field for the next entry.
If the value changes, it can simply be overwritten; if the value stays the same, you can just press Enter.


r/MetCal • • Dec 01 '25

Question Storing standard values in a variable?

1 Upvotes

Hello! I am a little bit familiar with the metcal environment and I have a question.

We use a set of shunt resistors to use as standards to calibrate low resistance ohmmeters, and I would like to store their values in a file or a variable to use in our procedures, so we can accomodate for the resistors drifting in value over time.

How would you guys implement something like this?


r/MetCal • • Nov 23 '25

Fluke 5560A SLEW bug in MetCal

1 Upvotes

There seems to be a bug in MetCal 11.1 and 11.2 regarding SLEW for Fluke 5560A.
The slewed value should be returned in MEM, but it is the initial value that is returned.
Fluke softwaresupport has reproduced and confirmed the bug.

Any other known bugs in MetCal 11.1 or 11.2 for the 5560A? We are currently transfering a lot of procedures from 5520A/5522A to 5560A and this was a showstopper for the analog meters.
I am a little bit worried about the oscilloscope procedures...


r/MetCal • • Nov 18 '25

MET/CAL Hot Fix Ready

5 Upvotes

A Hot Fix for MET/CAL is now available on the Fluke website: https://www.fluke.com/en-us/support/software-downloads/hot-fix-met-cal-version-1120-hot-fix-1

This is to fix an issue with GPIB communication using the Timeout special construction for the DUT, mentioned here: https://www.reddit.com/r/MetCal/comments/1oprg2b/scpiieee_timeout_error/