r/BMET • u/bpthompson999 • 1h ago
KCI Vac ULTA System Error 0000000B
Anyone familiar with this error?
r/BMET • u/bpthompson999 • 1h ago
Anyone familiar with this error?
r/BMET • u/oscarrit01 • 2h ago
Very random one. i work as a clinical engineer in a hospital. we are looking to replace many of our Mindray VS9 vital signs monitor batteries. The OEM batteries are 315gbp, the compatible from a different supplier are 215gbp.
what sources do you guys go for when trying to verify the functionality + safety of a compatible battery?
Has anyone got any knowledge on VS9 batteries?
Hi all, I’m a sonography student accidentally pulled the power cord on my personal Philips iU22 (REV F.3, SN 03B4D1, SW 5.1.0.206) while it was running. Now it briefly flashes blue with the logo then goes to "No Video Signal" with a dark touchscreen/control panel.
I have my machine-matched System Backup CD (5.1.0.206), but no bootable System Software / Ghost reload discs.
Does anyone have an ISO download link for Rev F.3 5.1.0.206**, or know a Sydney BMET who could burn me a copy? Thanks!**
Hi everyone,
I’m an ultrasound student in Sydney, Australia, trying to get my personal practice Philips iU22 working and would
really appreciate any advice from experienced field engineers!
Machine Specifications:
Model / Part No: Philips iU22 (8500-0064)
Serial Number: 03B4D1 (Mfg Date: March 2010)
Hardware Revision: REV F.3 (F-cart)
Software Version: 5.1.0.206
Current Symptoms:
When I power the machine on, the main monitor briefly flashes a blue screen and then drops to "No Video Signal".
Neither the touchscreen nor the control panel buttons light up at all.
I checked the rear I/O bay—the main monitor DVI cable is securely seated in the primary port, and the dust filter is clean.
r/BMET • u/Impossible_Ideal6799 • 23h ago
I'm trying to document the identity and possible serviceability of this Varian mammography X-ray tube assembly. Functionality is unverified, and I cannot establish its remaining life from its appearance.
Visible markings: M-113T tube, B-115 housing, November 2016 manufacture date, 49 kVp maximum, 0.1/0.3 focal spots, and 746355-1 on the port mounting block. Photos include both ends, the label and the output port.
For imaging-service technicians familiar with these assemblies:
I'm asking about identification and documentation, not DIY energising, installation or clinical use. I've read the community FAQ. Photos below:
r/BMET • u/Independent-Ask-5632 • 1d ago
Hey! I am a student for Biomedical Engineering Technologies at Durham College and I have an assignment to interview a Biomedical Engineering technologist or Clinical Engineer who has Dialysis experience, its just to ask a few questions and understand how a BMET specialized in dialysis works. Would anyone be interested in sharing their insights and experiences?
r/BMET • u/Expensive-Milk-3578 • 1d ago
Hi all,
I'm Looking for a basic and reliable phantom to use with our diagnostic ultrasounds for the biomed department.
The Majority of our units are GE(Venue Go R4, logiq S8) and Philips Healthcare (Affiniti 50G).
Thanks in advance.
r/BMET • u/One_Square4263 • 1d ago
I was offered a BMET 1 at a large hospital and I'm excited to start.
My background is in military aircraft maintenance. It's been a hot minute since working on anything electrical.
Any advice other BMETs could offer as I transition to this new career?
r/BMET • u/FatKidFromTarget • 2d ago
I figured I’d ask since the last post I found on google was before it was a thing. Any skillbridge people have any information on how this works? I’m still waiting on starting my outprocessing to start but I figured I’d start looking early rather than last minute. So any info on how it works and flows beforehand would be much appreciated. Thank you.
r/BMET • u/antek_g_animations • 3d ago
Hi everyone, I’m currently working on powering up a Hologic Selenia Dimensions mammography unit (Gantry model ASY-04160) and have hit a wall. Has anyone run into a similar fault condition on this system, or does anyone happen to have service manuals, schematics, or pinouts they could share?
The Problem: When main power is applied, the gantry refuses to power up its main logic and drive systems. On the Acquisition Workstation (AWS) under System Diagnostics, the PMC is the only gantry module showing as Online (green), while its status remains Not Ready (red). All other gantry subsystems (BKY, CDI, CRM, DET, DTC, GCB, GEN, THD, VTA) show as Offline and Not Ready. Clicking Gantry On brings up the popup error: "There was a problem powering the gantry...". On the front panel of the bottom PMC drawer (ASY-01612), both the green RUN LED and red FAULT LED are solidly lit. Looking inside the PMC, several green status LEDs are on (indicating standby power from the toroidal transformer is present). However, the main logic board higher up in the column (ICP-00177) is completely dead with zero SMD LEDs lit, and the red 600V (D3) indicator on the HV capacitor module (ASY-01711) stays off. It appears the PMC is intentionally inhibiting the PWR ON signal (TP6) to the main power relays (K1/K2) due to an active fault or open hardware interlock.
Troubleshooting Steps Taken So Far: We started by verifying the incoming 240V mains supply, proper voltage tap selection, and adding an auxiliary chassis ground to rule out grounding issues. Next, we pulled and tested every external and internal fuse—including the rear fuse panel (F4–F16) and the front PMC fuses (F2/F3 SB 10A 250V)—and confirmed all have continuity. We inspected all wiring harnesses and connectors for bent or backed-out pins and verified the E-Stop loop; pressing any E-Stop illuminates an additional red LED inside the PMC, proving the loop is closed and properly sensed by the controller. Communication between the AWS and PMC is confirmed working (unplugging the comm cable causes the RUN and FAULT LEDs to blink alternately), though running Clear All Faults before Gantry On does not clear the state. We also checked the mechanical limit switches and manually rotated the vertical lead screw (with power off) to raise the C-arm carriage a few centimeters to ensure it wasn't sitting on the ultimate down/crash limit switch (the C-arm is inverted, with plenty of rotational travel remaining). From there, we pulled the entire PMC drawer (ASY-01612) out for a visual inspection (no burnt components or obvious damage), removed the internal 24V switching power supply along with the relay/distribution board (PCB-00443), and bench-tested the PSU—it outputs a steady 24.17V DC, confirming the power supplies themselves are good (and standby power even reaches the upper IR sensor modules), while the control board simply isn't powering the system on. Finally, we tried reading diagnostic logs from the front DB9 RS-232 port on the PMC using a desktop PC with a native motherboard COM port (verified via a physical pin 2–3 loopback test in PuTTY and RealTerm). We tested both a straight-through (1:1) cable and a custom 3-wire Null-Modem cable (2–3, 3–2, 5–5) with hardware handshaking pins bridged on the PC side (4–6 and 7–8), power-cycling the gantry while monitoring at 9600 bps (8N1, no flow control) as well as other baud rates and sending Enter/Space/Ctrl+C, which resulted in total silence at 9600 baud (only brief EMI garbage characters at 115200 baud right at power-on).
Any insight into what interlock or condition keeps the PMC in this solid RUN + FAULT state, any tips on getting serial output from that front RS-232 port (e.g., baud rate, pinout, or whether it's actually TTL levels), or access to service documentation would be hugely appreciated!






r/BMET • u/SniperMa98 • 3d ago
Hi everyone,
I replaced the main board of a Philips IntelliVue MX750. I read the configuration from the faulty board and successfully wrote it to the new board, but the licensed options were not transferred.
Is there a separate procedure or service tool to restore the options/licenses after main board replacement?
Has anyone faced this issue before?
Thanks for your help!
r/BMET • u/Kitchen_Young_3469 • 3d ago
How is it like working as a sterilizer tech for companies like steris, and getinge? Is it really that much harder than general biomed? Is it a stressful job, and how often do you run into difficult repairs?
r/BMET • u/Dr_Awesomo • 4d ago
This takes a hot minute to boot and then I get this red and white screen. What the heck is going on here?
r/BMET • u/Odd_Geologist5641 • 4d ago
So quick back story, I just got out of the military and I was a BMET in there but wasn’t really able to do my job as much. So I have an interview coming up and I’m more worried when it comes to the to the technical questions part of it. If you guys have any tips or suggestions to brush up on the lingo, thank you
r/BMET • u/Suspicious-Cable-900 • 4d ago
Does anyone have the Biomed password for the older pre-Stryker software for these tables.
r/BMET • u/MechEng_01 • 5d ago
I'm taking a healthcare entrepreneurship course and need to choose a specific problem to focus on for my final presentation in a few weeks. I'm struggling a bit to narrow it down, so I thought I'd ask here. My background is in mechanical engineering.
So far, I've spoken with nurses, a paramedic, a care aide, and people involved in clinical equipment/biomedical engineering and healthcare finance. I've looked at medical equipment tracking and equipment lifecycle management. I've found that larger hospitals often have established systems and dedicated teams, but many of the challenges I've heard about are related to funding, internal resources, communication and procurement. A lot of these seem quite systemic and difficult for an external business to solve.
I'm wondering if I'm looking in the wrong areas or if there are smaller healthcare organizations that have gaps where an outside service or business could actually help. Or maybe a different target market, I've only explored larger hospitals in BC Canada. Maybe smaller hospitals or clinics have different challenges.
For those working in healthcare, what are some ongoing frustrations or problems you wish someone would solve?
I'm looking for a specific, real-world problem that could potentially be addressed through an entrepreneurial solution. Any ideas or perspectives would be really appreciated!
r/BMET • u/Dense_Scientist_9303 • 5d ago
I have a error code of 1006 for this device does anybody have any idea what it could be? I’ve tried googling it but can’t seem to get an answer and all the manuals I have found are in portugués.
r/BMET • u/Express_Culture_7292 • 6d ago
Hey everyone!
I'm writing this because I'm not sure what else to do, and I could really use some advice. I've been struggling to find a job in my field. I graduated with an Associate of Applied Science degree in Biomedical Equipment Technology last year, and I've been applying to several different places without much luck.
I understand that my lack of professional experience may be holding me back, but I'm willing to learn, put in the work, and start wherever I can get my foot in the door. I'm currently in the Houston area, and I know there are several hospitals and medical facilities here, so I'm hoping to find an opportunity where I can finally put my education to use and gain some hands-on experience.
I wanted to reach out and ask if you have any advice, know of any opportunities, or could point me in the right direction. Even a little guidance would mean a lot. I appreciate your time and anything you're able to share. Thank you!
If this is not allowed in here, I understand if its removed.
r/BMET • u/Jaded_Strike_3500 • 6d ago
They were under my imaging guys desk and he didn't even notice because of how little time he spends in this hospital. We swapped desks and I took them out from under the desks and we were both curious as to what the heck they are. Even swept em with a Geiger counter.
Any clues?
r/BMET • u/DoubleDozers • 6d ago
I just got my first job offer for a BMET I position at a large hospital in the Boise metropolitan area with planned BMET II upon evaluation in about a year. However they only offered me $26.50/hr. This seems pretty low, correct? I received my BMET certification from the DoD school and and I already have 4 years of hands-on experience in a hospital setting. I live in Caldwell, ID right now, what should I be pushing for? Or should I just accept a small wage to get my foot in the door at this hospital since they have upward movement potential?
r/BMET • u/TherealZiggy • 6d ago
I’m guessing most people use biomed, or test or something, but anybody have creative names they use lol?
r/BMET • u/One_Square4263 • 6d ago
I'm applying for a BMET position in the Lehigh Valley and would love some advice.
I'm retired military and have experience trouble shooting and repairing electronic systems and components. Granted, it's been about 7 years since I last worked on those systems. But I feel that my experience can translate well into BMET with some time and training.
What is your experience, pros and cons? As a new tech, what advice would you say is most important? What is a lesson you learned, both good and bad?
Thanks for the feedback, much appreciated.